SM731_K2200_01_en 06/2009
6-2
Rated high voltage and maximum corresponding current
70 kV, 7 mA
Current/voltage combinations for a maximum output power of:
490 W, 70 kV / 7 mA
Rated power for exposure time of 0.1 sec.
490 W
Rate of use
At 70 kV, 7 mA and 0.1 sec. and at the maximum tank temperature: one
exposure every 8 sec.
Minimum value of the current/time product in the range of
conformity
0.07 mAs at 7 Ma
Selection of parameters
70 kV / 7 mA and 60 kV / 7 mA
Area of conformity to IEC standard 60601-2-7
Reproducibility of the emitted radiation: conform
Linearity of the emitted radiation: conform
Precision in radiography: conform
Measurement conditions
kV: Indirect measurement using a kV peakmeter
mAs: Direct measurement in the circuit using a mAs-meter
Exposure time: Indirect measurement on the kV signal at 75% of the
peak value
Type of protection against
electric shock
Class 1 equipment
Degree of protection against
electric shock
Type B
Protection against harmful
ingress of water
Ordinary equipment
Operation mode
Continuous operation with
intermittent loading
Flammable anesthetics
Not suitable for use in presence of
flammable anesthetics or a
mixture of flammable anesthetics
with air or oxygen or nitrous oxide