X-Ray Dose Emission Information
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02 5–19
X-Ray Dose Emission Information
Compliance with EURATOM 97/43 Directive
You can right-click on each image to display the estimated emitted dose received by the
patient. You can use this information to calculate the effective dose received by the
patient for the image.
The radiation emission dose is expressed in mGy.cm2. This dose is measured at the
primary collimator outlet. The dose is accurate to +/-30%. The primary slot is 0.5 mm
wide and 18.2 mm high.