PREVENTIVE MAINTENANCE
10DEC05
PM4825-1
Page
11 of 30
Cleaning the BAR CODE READER
1 If necessary, remove the CASSETTE
HANDLING AY.
Caution
Use the VACUUM CLEANER to remove
dust before using the CANNED AIR.
2 Use the VACUUM CLEANER to remove
dust from the BAR CODE READER.
3 If necessary, use the CANNED AIR to
remove the dust from the BAR CODE
READER.
4 If the CANNED AIR does not remove the contamination from the BAR CODE READER,
clean the BAR CODE READER with:
LENS CLEANER SOLUTION 1C8000
LENS PAPER 952807
5 Continue with Cleaning the EXTRACTION BAR.
H194_1080AC
READER
BAR CODE
H194_1080ACA