Installing the New LIGHT SEALS on the TOP COVER
1 Open the TOP COVER.
Important
The narrow end of the LIGHT SEAL must be toward the front edge of the SLOT.
The LIGHT SEAL must make a seal against the back edge of the SLOT.
2 Remove:
LEFT LIGHT SEAL - discard
RIGHT LIGHT SEAL - discard
3 Install the new:
LEFT LIGHT SEAL
RIGHT LIGHT SEAL
4 Continue with Installing the New CLAMP BAR PLUSH.
Top View
H194_1083BC
SEAL
LIGHT
RIGHT
SEAL
LIGHT
LEFT
H194_1083BCA
TOP COVER
narrow ends of
LIGHT SEALS
back edge of SLOT
front edge of SLOT