Operating Instructions OIO 14en Edition 07/2017 45
Troubleshooting
Troubleshooting
Tab. 1 Fault table
16 Outgassing Increase the system pressure.
Reduce the temperature.
17 Pulsations too high Use another feed pump.
Carry out changes to the system.
18 Back pressure too low Increase the back pressure.
19 Flow rate fluctuations too high Ensure a continuous flow rate by taking suitable
measures (use of a different pump. valve, damper,
etc.)
- or -
Smoothen the indication,
while observing the electronic operating
instructions.
20 Filling amount too low Use a suitable flowmeter size.
Use a suitable flowmeter type.
21 Strongly deviating operating data Use a suitable flowmeter type.
Adapt the operating data to the flowmeter.
22 Wear at the measuring unit and
bearing
Renew the measuring unit.
Replace the bearing.
Filter out the abrasive materials.
23 Sluggishness through deposits Disassemble the flowmeter and clean it carefully.
24 Flow impaired at the system end Check whether the fluid flows in the system (pump in
operation, slide valve opened, etc.).
Check whether shut-off devices before and after the
flowmeter are opened.
25 Flowmeter switched to bypass Switch the flowmeter to through-flow.