Lake Shore Model 331 Temperature Controller User’s Manual
MNMX Minimum and Maximum Input Function Parameter Command
Input: MNMX <input>, <source>[term]
Format:
a,n
<input> Specifies input to configure: A or B.
<source> Specifies input data to process through max/min. Valid entries: 1 = kelvin,
2 = Celsius, 3 = sensor units, 4 = linear data.
Example: MNMX B,3[term] – Input B min/max function is on and processes data from the input sensor
units reading.
MNMX? Minimum and Maximum Input Function Parameter Query
Input: MNMX? <input>[term]
Format:
a
<input> Specifies which input to query: A or B.
Returned: <source>[term]
Format: n (Refer to command for description)
MNMXRST Minimum and Maximum Function Reset Command
Input: MNMXRST[term]
Remarks:
Resets the minimum and maximum data for all inputs.
MODE Remote Interface Mode Command
Input: MODE <mode>[term]
Format:
n
<mode> 0 = local, 1 = remote, 2 = remote with local lockout.
Example: MODE 2[term] – Places the Model 331 into remote mode with local lockout.
MODE? Remote Interface Mode Query
Input: MODE?[term]
Returned:
<mode>[term]
Format: n (Refer to command for description)
MOUT Control Loop Manual Heater Power (MHP) Output Command
Input: MOUT <loop>, <value>[term]
Format:
n,±nnnnnn[term]
<loop> Specifies loop to configure: 1 or 2.
<value> Specifies value for manual output.
Example: MOUT 1,22.45[term] – Control Loop 1 manual heater power output is 22.45%.
MOUT? Control Loop Manual Heater Power (MHP) Output Query
Input: MOUT? <loop>[term]
Format:
n
<loop> Specifies which loop to query: 1 or 2.
Returned: <value>
Format: ±nnnnnn[term] (Refer to command for description)
Remote Operation 6-35