EasyManua.ls Logo

Lakeshore 331S - 4.16.2 Analog Output in Manual Mode

Lakeshore 331S
182 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lake Shore Model 331 Temperature Controller User’s Manual
4.16.2 Analog Output In Manual Mode
In Manual mode, the analog output provide a fixed output according to a percentage of full scale
entered by the user; -100.00% to +100.00% corresponding to –10 V to +10 V. The setting resolution
on the display is 0.001% but the output itself is limited to 0.003%. Press the
Analog Output key.
Select With °®
Analog Out Manual
Press the s or t key until “Manual” is showing. Press the Enter key.
Select for AnOut °®
Bipolar Mode On
Press the s or t key to toggle between Bipolar Mode On or Off. Bipolar mode refers to whether or
not negative voltages are used, as shown below.
Bipolar Mode: On
Bipolar Mode: Off
10 V
+10 V
0 V
100%
User Entry
Output
+10 V
Output
Manual Mode
+100%
0%
0 V
+100%
0%
User Entry
For this example, we will choose Bipolar Mode On. Press the Enter key.
Enter for AnOut
Manual Out - 50%
The desired fixed output you want as a percent of full scale is entered using the numeric keypad,
which includes the numbers 0
9, +/–, and decimal point. For this example, we will enter –50%. Press
the
Enter key. The instrument will return to the normal display. The analog output will begin to output
a constant voltage that is –50% × 10 volts = –5 volts.
In a second example, if you repeat the same procedure using all the same settings, but enter +75%,
then the output would be +75% × 10 volts = +7.5 volts.
In a third example, if you repeat the same procedure, but choose Bipolar Off, enter +25%, then the
output would be +25% × 10 volts = +2.5 volts. The difference being that any negative sign will be
ignored with Bipolar Mode Off and the output will always be a positive voltage.
4-34 Operation

Table of Contents

Related product manuals