EasyManua.ls Logo

Lakeshore 332 - Page 16

Lakeshore 332
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lake Shore Model 332 Temperature Controller User’s Manual
Table 1-2. Model 332 Typical Sensor Performance Chart (Continued)
Sensor Type
Germanium
GR-200A-1500
Germanium
GR-200A-250
Cernox™
CX-1070
Cernox™
CX-1050
Cernox™
CX-1030
Temperature Coefficient Negative Negative Negative Negative Negative
Input Range AutoRange AutoRange AutoRange AutoRange AutoRange
Sensor Excitation*
(Constant Current)
75 mV max
4 ranges from 75 Ω -
75 kΩ
75 mV max
4 ranges from 75 Ω -
75 kΩ
75 mV max
4 ranges from 75 Ω -
75 kΩ
75 mV max
4 ranges from 75 Ω -
75 kΩ
75 mV max
4 ranges from 75 Ω -
75 kΩ
Display Resolution (Sensor Units) 5 digits 5 digits 5 digits 5 digits 5 digits
Example Lake Shore Sensor
GR-200A-1500
with1.4D calibration
GR-200A-250 with
0.3D calibration
CX-1070-SD with
1.4L calibration
CX-1050-SD with
1.4L calibration
CX-1030-SD with
1.4L calibration
Standard Sensor Curve
Requires calibrated
sensor
Requires calibrated
sensor
Requires calibrated
sensor
Requires calibrated
sensor
Requires calibrated
sensor
Typical Sensor Sensitivity
–64200 Ω/K at 1.4 K
–668 Ω/K at 4.2 K
–0.078Ω/K at 77 K
–8450 Ω/K at 1 K
–68.9 Ω/K at 4.2 K
–0.054 Ω/K at 77 K
–17600 Ω/K at 4.2 K
–-969 Ω/K at 10 K
–8.26 Ω/K at 77 K
–0.419 Ω/K at 300 K
–-42800 Ω/K at 2 K
–2290 Ω/K at 4.2 K
–2.15 Ω/K at 77 K
–0.131 Ω/K at 300 K
–-8670 Ω/K at 1.4 K
–138 Ω/K at 4.2 K
–-.828 Ω/K at 77 K
–0.067 Ω/K at 300 K
Measurement Resolution:
Sensor Units
Temperature Equivalence
A
uto Range
see Table 1-3
<10 µK at 1.4 K
4
30 µK at 4.2 K
3
3.8 mK at 77 K
1
A
uto Range
see Table 1-3
<10 µK at 1 K
3
40 µK at 4.2 K
2
5.5 mK at 77 K
1
A
uto Range
see Table 1-3
<10 µK at 4.2 K
4
<100 µK at 10 K
3
3.6 mK at 77 K
2
7.2 mK at 300 K
2
A
uto Range
see Table 1-3
<10 µK at 2 K
4
<10 µK at 4.2 K
3
1.2 mK at 77 K
2
2.3 mK at 300 K
1
A
uto Range
see Table 1-3
<10 µK at 1.4 K
3
20 µK at 4.2 K
2
3.6 mK at 77 K
2
4.5 mK at 300 K
1
Electronic Accuracy:
Sensor Units
Temperature Equivalents
Auto Range
see Table 1-3
±0.2 mK at 1.4 K
4
±1 mK at 4.2 K
3
±38 mK at 77 K
1
Auto Range
see Table 1-3
±0.2 mK at 1 K
3
±2 mK at 4.2 K
2
±47 mK at 77 K
1
Auto Range
see Table 1-3
±1 mK at 4.2 K
4
±3 mK at 10 K
3
±28 mK at 77 K
2
±128 mK at 300 K
2
Auto Range
see Table 1-3
±0.3 mK at 2 K
4
±1 mK at 4.2 K
3
±30 mK at 77 K
2
±130 mK at 300 K
1
Auto Range
see Table 1-3
±0.2 mK at 1 K
3
±2 mK at 4.2 K
2
±57 mK at 77 K
2
±224 mK at 300 K
1
Temperature Accuracy including
electronic accuracy, CalCurve
and calibrated sensor
±6 mK at 1.4 K
4
±6 mK at 4.2 K
3
±128 mK at 77 K
1
±6 mK at 1 K
3
±7 mK at 4.2 K
2
±137 mK at 77 K
1
±7 mK at 4.2 K
4
±11 mK at 10 K
3
±78 mK at 77 K
2
±268 mK at 300 K
2
±6 mK at 2 K
4
±7 mK at 4.2 K
3
±80 mK at 77 K
2
±270 mK at 300 K
1
±6 mK at 1.4 K
3
±8 mK at 4.2 K
2
±107 mK at 77 K
2
±364 mK at 300 K
1
Control Stability:
Sensor Units
Temperature Equivalence
±80 mΩ
±20 µK at 1.4 K
4
±60 µK at 4.2 K
3
±7.6 mK at 77 K
1
±80 mΩ
±20 µK at 1 K
3
±80 µK at 4.2 K
2
±11 mK at 77 K
1
±80 mΩ
±20 mK at 4.2 K
4
±7.2 mK at 77 K
2
±14.4 mK at 300 K
2
±80 mΩ
±8 µK at 2 K
4
±20 µK at 4.2 K
3
±2.4 mK at 77 K
2
±4.58 mK at 300 K
1
±80 mΩ
±20 µK at 1.4 K
3
±40 µK at 4.2 K
3
±7.2 mK at 77 K
3
±9 mK at 300 K
3
Magnetic Field Use
Not Recommended Not Recommended
Recommended for
T > 2 K & B 19 T
Recommended for
T > 2 K & B 19 T
Recommended for
T > 2 K & B 19 T
* Current reversal eliminates thermal EMF voltage errors for resistor sensors.
Typical sensor sensitivities were taken from representative calibrations for the sensor listed.
NOTES:
1 NTC RTD Range 75 Ω
2 NTC RTD Range 750 Ω
3 NTC RTD Range 7500 Ω
4 NTC RTD Range 75000 Ω
1-6 Introduction

Table of Contents

Related product manuals