Lake Shore Model 332 Temperature Controller User’s Manual
Interface (Continued)
Select With °®
Emulation Mode 332
Use the s or t key to toggle between 332 and 330. The default setting is 332. If 332 mode is selected,
pressing the
Enter key will return you to the normal display.
To support owners of the Lake Shore Model 330 Temperature Controller, 330 Emulation Mode is
provided. The 330 Emulation Mode only affects remote operation; front panel operation of the Model
332 is not changed. In 330 Emulation Mode, curve locations are mapped to match Model 330 locations.
For example, the DT-500-D Curve, found at curve location 3 in the Model 332, is mapped to location 0
when in 330 mode. This applies to the following remote commands: ACUR, ACUR?, BCUR, BCUR?.
The following Model 330 commands are not supported in 330 Emulation Mode: CUID?, CURV, CURV?,
ECUR, KCUR, and SCAL. Please refer to your Model 330 User’ s Manual for any additional questions
concerning remote commands.
Selecting 330 Emulation Mode causes two additional screens to appear.
Select With °®
Sample Channel A
Use the s or t key to toggle between Sample Channel A or B. The default setting is A. Press the
Enter key. You will see the following display.
Select With °®
Sample Units K
Use the s or t key to cycle through the Sample Units: K, C, and Sensor, where K = kelvin,
C = degrees Celsius, and Sensor = volts (V) or ohms (Ω). The default setting is K. Press the
Enter key.
You will return to the normal display.
4.21 DEFAULT VALUES
It is sometimes necessary to reset instrument parameter values or clear out the contents of curve
memory. Both are all stored in nonvolatile memory called NOVRAM but they can be cleared
individually. Instrument calibration is not affected except for Room Temperature Calibration which
should be redone after parameters are set to default values or any time the thermocouple curve is
changed.
To reset the Model 332 parameters to factory default values, press and hold the Escape key until the
screen shown below appears.
Code Date: 06/01/00
Default Values Yes
Use the s or t key to select Yes or No to reset the NOVRAM. Select Yes to reset all Model 332
parameters to the defaults listed in Table 4-5. Press the
Enter key. The second screen appears.