EasyManua.ls Logo

Lakeshore 332 - Page 19

Lakeshore 332
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lake Shore Model 332 Temperature Controller User’s Manual
Specifications (Continued)
Control
Control Loops: 2
Control Type: Closed Loop Digital PID with Manual Heater Power Output, or Open Loop
Tuning: AutoTune (one loop at a time), Manual PID, Zones
Control Stability: Sensor dependent, refer to performance chart
PID Control Settings:
Proportional (Gain): 0
1000 with 0.1 setting resolution
Integral (Reset): 1
1000 (1000/s) with 0.1 setting resolution
Derivative (Rate): 1
200% with 1% resolution
Manual Heater Power: 0
100% with 0.001% setting resolution
Zone Control: 10 temperature zones with P, I, D, Manual Heater Power Output, and Heater Range
Setpoint Ramping: 0.1 to 100 K/min
Protection: Curve Temperature limits, Power up heater off, Short-circuit protection
Loop 1 Loop 2
Heater Output Type Variable DC current source Variable DC voltage source
Heater Output D/A Resolution 18 bit 16 bit
Max Heater Power 50 W 10 W
Max Heater Output Current 1 A 1 A
Heater Output Compliance 50 V 10 V
Heater Source Impedance N/A
0.1 Ω maximum
Heater Output Ranges 3 decade steps in power 1
Heater Load Type Resistive Resistive
Heater Load Range
10 Ω to 100 Ω recommended 10 Ω minimum
Heater Load for Max Power
50 Ω 10 Ω
Heater Noise (<1 kHz.) RMS
50 µV +0.017% of output voltage
<0.3 mV
Isolation
Optical isolation between output and
other circuits
None
Heater Connector Dual banana Detachable terminal block
Loop 1 Full Scale Heater Power at Typical Resistance
Heater Resistance Heater Range Heater Power
10 Ω
Low
Med
High
100 mW
1 W
10 W
25 Ω
Low
Med
High
250 mW
2.5 W
25 W
50 Ω
Low
Med
High
500 mW
5 W
50 W
Front Panel
Display: 2 line by 20 character, 9 mm character height, vacuum fluorescent display
Number of reading displays: 1 to 4
Display Units: K, °C, V, mV, Ω
Display Source: Temperature, sensor units, max, min, and linear equation
Display Update Rate: All readings twice per second
Temperature Display Resolution: 0.001° between 0°
99.999°, 0.01° between 100° 999.99°, 0.1° above 1000°
Sensor Units Display Resolution: Sensor dependent, to 5 digits
Other Displays: Setpoint, Heater Range, and Heater Output (user selected)
Setpoint Setting Resolution: Same as display resolution (actual resolution is sensor dependent)
Heater Output Display: Numeric or graphical display in percent of full scale for power or current
Heater Output Resolution: 1% numeric or 2% graphical
Display Annunciators: Control Input, Remote, Alarm, Tuning, Ramp, Max, Min, Linear
Keypad: 20 full-travel keys, numeric and specific functions
Front Panel Features: Front panel curve entry, display brightness control, keypad lock-out
Introduction 1-9

Table of Contents

Related product manuals