Overview of TDS 6 Set-Up Menu
Select Cell Constant
It is possible to select a cell constant of K = 1.0, 10, or 0.1.
Use a cell of K = 1.0 for midrange measurements
Use a cell ofK=10forhigh range measurements
(above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range measurements
(below 20 µS or 10 ppm).
The cell included with the meter has a cell constant
of K = 1.0.
1. Enter Advanced Set up.
2. Press or button until [CELC] appears on the
LCD. Press ENTER button.
3. Press or button to select either “1.0”, “0.1” or
“10.0”. Ensure the cell constant selected correspond
with the conductivity electrode that is being used
with the meter.
4. Press ENTER button to select. The meter will
return to the menu, [CELC].
5. Press or button to move to the next menu or
press CAL to exit to measurement mode.
%
C°
Meter Off
Press ON/OFF and MODE buttons keys
simultaneously for 2 seconds, release
Release ON/OFF button first then release MODE button
a second later.
A "StUP" indicator will appear for 1.5
seconds before showing the first menu.
ON
OFF
MODE
HOLD
ENTER
25

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LaMotte 6 Series and is the answer not in the manual?

LaMotte 6 Series Specifications

General IconGeneral
BrandLaMotte
Model6 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals