09/2021 ECO Gold 109 / 134
- Access to the main menu level is obtained by pressing the enter key
.
- Selection and confirmation of Setup Control
Control parameter extern Pt100.
The adjacent menu window appears. Apart from the control parameters the
currently set values are displayed.
- Under the menu point "Tv manual/auto" you can select between man-
ual and automatic entry using .
The selection "automatic" is displayed in the menu line by (auto). If "auto-
matic" is selected, the entry is blocked for the parameters Tv and Td.
- Select and confirm parameters with or and .
The respective edit window appears with Min and Max figures for the pa-
rameter values Kpe, Tne, Tve, Tde and Xpf.
- Change the value with or .
- Single figures can be selected by pressing or .
- Confirm your choice with the enter key .
-
By pressing (ESC) you are returned to the menu level without any change.
18.2.1 Setting the correcting quantity limit
- Access to the main menu level is obtained by pressing the enter key
.
- Selection and confirmation of Setup Control.
The adjacent menu window appears.
- Select and confirm Correction limit. with or and .
The adjacent menu window appears. The minimum and maximum possible
values and the current value are displayed.
- Change the value with or .
- Single figures can be selected by pressing or .
- Confirm your choice with the enter key .
-
By pressing (ESC) you are returned to the menu level without any change.