30 / 134 ECO Gold 09/2021
6.2 Connection of external consumers
For heating thermostats a pump connection set is available as an accessory ( 9) for the connection of an external
consumer.
This pump connection set is included as standard with cooling thermostats and with the heating thermostats E 4G and
ET 15 G.
Notice
Confusing pump connector and cooling coil
Environmental hazard from leaking heat transfer liquid
Follow the illustrations in this section.
Notice
Leaks from consumers, hoses and accessories
Environmental hazard from leaking heat transfer liquid
Always secure the hoses with suitable safety devices.
The ECO thermostat can be equipped as an immersion thermostat or as a circulation thermostat.
Immersion thermostat/heating thermostat
With heating thermostats the control head must first be removed by releasing the two cross-head screws from the
bath bridge.
For optional operation with the pump first mount the pump connection set and then carry out the complete assembly:
Cut the thread with the screw
- Cut the thread on the holed flange already before assembly.
The pump connection set can be mounted on one side of the control
head (see illustration).
- Withdraw the mains plug.
- Use a soft underlay to avoid scratches to the upper side of the
control head.
- With heating thermostats: Remove the flat seal.
- Remove the blind flange by releasing the two cross-head screws.

Table of Contents