09/2021 ECO Gold 51 / 134
7.4.4 Activating the "Standby" operating state
In the "Standby" mode the pump, heater and chiller are switched off. The operating display remains active.
- Activate Standby by pressing (right soft key).
7.4.5 Defining temperature limits
With this function the temperature limits Til and Tih are defined. If, for example, you are using water as the heat transfer
liquid, +5 °C is practicable as the minimum temperature and +95 °C as the maximum temperature.
- Access to the main menu level is obtained by pressing the enter
key .
- Selection and confirmation of Setup Limits.
The adjacent menu window appears.
- Select the lower (Til) or upper (Tih) limit with or and con-
firm it with .
In the entry window the cursor flashes below the value to be changed.
The permissible adjustment range is indicated with Min and Max.
- Change the value with or .
- Single figures can be selected by pressing or .
- By pressing (+/-) the arithmetic sign can be changed.
- Confirm your choice with the enter key .
-
By pressing (ESC) you are returned to the menu level without any change.