6. Information on specific controls
The ACC (Attendant Control Compact) is fitted to advanced hospital and patient care beds for use where
patient positioning must be carefully controlled by medical staff.
1. ACC (MEDLINE
®
CARELINE
®
)
2. ACK (MEDLINE
®
CARELINE
®
)
With the OpenBus
TM
system it is possible to use keypads as attendant control or hand controls integrated in the bed side rails.
There are two different variants of ACK: ACK1 and ACK3.
ACK1 is with a single keypad, whereas ACK3 is able to work with two keypads at the same time (typically an inside side rail and an outside side rail).
Features and Options:
Straight cables: 1250 mm, 1800 mm or 2500 mm
The standard ACK colour is grey (RAL 7035)
Usage:
Operation temperature: +5 °C to +40 °C
Storage temperature: -10 °C to +50 °C
Compatibility: LINAK OpenBus™ control boxes
Relative humidity: 20 % to 80 % – not condensing
Atmospheric pressure: 700 to 1060 hPa
Meters above sea level: Max. 2000 meters
Approvals: IEC60601-1
ANSI/AAMI ES60601-1
CAN/CSA-22.2 No 60601-1
Generel information
For LINAK standard ACK’s, the following is applicable:
Adhesive for the standard ACK is 3M 468MP
For information re. suitable and unsuitable surfaces, please see 3M’s
webpage
Standard recommandation for curing time is 72 hours
The customer is responsible for correct mounting on suitable surfaces
Recommendations
The customer responsibility includes making a proper design of the cable strain relief inside the side rail panel.
The customer should consider and ensure to use the correct flammability classification when defining and specifying the housing.
According to IEC 60601-1. Minimum HB plastic.
The customer should consider the existence of vibrations when defining and specifying the housing, i.e. we recommend the customer to carry
out a vibration test on the final product.
The customer must ensure a proper IP rating/test
The customer must ensure proper drop testing according to IEC60601-2-52 §201.15.3.4.1. In this clause there is an additional reference to
IEC60601-2-31.
The customer is responsible for correct mounting of the PCBA. Among other things, it means
- ensuring proper and safe mounting of the PCBA into e.g. the side rail
- ensuring proper and correct mounting between key pad connection tails and the ACK PCBA
- ensuring proper and correct mounting of the key pad
- the customer should consider proper precautions against ESD (Electrostatic discharge).
When handling ESDS (Electrostatic Discharge Sensitive) devices – e.g. during transport, storage, handling, production or mounting in an
application - exposure to harmful ESD must be avoided.
Consider proper creepage and clearance measures to fulfil IEC 60601. With One MOPP (One Means Of Patient Protection / Secondary
side of the actuator system)
It is not recommended to dismount the keypad after mounting as this may cause damage.