D9U001PPB-010134/106
7.6 Electromagnetic compatibility
Bed is intended for hospitals except for near active HF surgical equipment and the RF shield-
ed room of a medical system for magnetic resonance imaging, where the intensity of EM
disturbances is high.
Bed has defined no essential performance
List of used cables:
► Mains cable, maximum length 6 m
WARNING!
It is recommended to avoid the use of this device next to or in block with other
device, because it could lead to improper operation. If such use is needed, this
device and the other equipment should be under surveillance to verify proper
operation.
WARNING!
Use of the accessories, converters and other cables, than specified and
provided by manufacturer of this bed could lead to increase of electromagnetic
emission or lower the electromagnetic immunity of this bed and lead to improper
operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables
and external antenna) should not be used closer than 30 cm (12 inches) from any
part of this bed AVE2, including cables specified by manufacturer. Otherwise this
could lead to deterioration of functionality of this bed.
WARNING!
Do not overload the bed (SWL), respect the duty cycle (INT.) Maintenance in order
to maintain the basic safety with regard to electromagnetic disturbances for the
expected service life.
7.6.1 Manufacturer instructions – electromagnetic emissions
EMISSION TEST COMPLIANCE
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Complies