LEGENDA
N Normal movement: 12 s rotation and 3 s rest phase
a
The basic washing temperature refers to the temperature at which the heating is
switched off
b
The bath level is measured from the bottom of the tank after 1 min. of washing
machine operation and 30 s of the rest phase
d A tolerance of 20 s is allowed for the rest phases
e
No movement during heating until it reaches the set temperature -5 ° C. From the
set temperature -5 ° C until the set temperature is reached, a slight movement is
performed
f
Cooling: topping up with cold water to a bath level of 130 mm and operation for
another 2 min
g The rinsing time is measured after reaching the bath level
h
Heating to 40 ° C, maintained for 15 min. while operating before heating to the
washing temperature
11.3.3 Recommended chemical concentrations for disinfection the top cover
WARNING!
Do not use disinfectants with active substance: alcohol over 70 %, active chlorine
over 10,000 ppm, iodine, aldehydes and phenols
Active
substance
Concentra-
tion
pH range Dwell time
Post
application
Recommendations for use
Chlorine and
compounds
1 000 -
10,000 ppm
7 - 9 <10 mins
Strength and
microbial
target
dependent
Rinse and dry Acceptable and recommended
1000ppm is appropriate for day
to day cleaning. For routine
decontamination, use a concentration
of 5,000 ppm chlorine, in case of
heavy contamination, a concentration
of up to 10,000 ppm chlorine can be
used.
Must be rinsed with clean water and
towel dried (avoid harsh abrasion).
Alcohol 70 % 10 mins (max) Rinse and dry Acceptable for use and
recommended subject to
precautions
Coating may swell and be prone to
damage. Care must be taken until
fully dry (~1hr).
Glutaraldehyde 2 % 7.5 - 8.5 Temperature
dependent
Rinse and dry Acceptable for use
Ethylene Oxide 3 % >30 mins
Temperature
dependent
Rinse
thoroughly and
dry before use
Acceptable for sterilization only
Ensure that appropriate sterilization
validation has been completed.
Requires cleaning afterwards.