D9U001PPB-010138/106
8.1 Transport, braking
The birthing bed also allows for the transport of a patient. This is, however, subject to the
following conditions:
WARNING!
The birthing bed allows for the transport of a patient by pushing forward using
the handle on the back section. Before transporting the patient, disconnect and
store the power supply cables and equipotential and other connections properly,
to prevent their tearing off. Use the handle on the back section only to transport
the bed. Do not transport the patient if the bed is set in the chair position. If
you do so, you can cause injury to persons or damage to the device. Before
transporting the patient, the central brake must be released
WARNING!
Before running the device ensure that the front wheels are set in the direction of
travel, meaning that the front wheels are not pushed by the bed. In this situation
the wheels can be damaged.
WARNING!
When transporting the patient, the locking pedal must be used to secure the
front directional castor in a straight direction. Otherwise the patient has to be
transported by two persons.
WARNING!
When transporting the bed within hospital premises, make sure that the patient
as well as all equipment, such as cables (including the power supply cable),
hoses, tubes, etc., are safely placed in the area bounded by the circumference of
the bed. Otherwise you can cause injury to persons or damage to the device.
WARNING!
During transport, check the stability of the infusion stand - it can hit doors or
ceiling lights. Before transporting the patient, set the stands to the lowest position.
Otherwise you can cause injury to persons or damage to the device.
WARNING!
After the transport, reconnect the power supply cable and secure the bed with
the central brake. Otherwise you can cause injury to persons or damage to the
device.
WARNING!
Do not expose the bed to prolonged extreme effects of external environment
(snow, frost, rain, etc.). It could damage the device.
WARNING!
Transport the bed in its lowest position and make sure that it is not transported
over areas with lateral inclination exceeding 5°. In case of interrupting the
transport (stopping the movement of the bed), always secure the bed with the
central brake. Otherwise you can cause injury to persons or damage to the
device.
WARNING!
The maximum threshold crossing height is 20 mm.