EasyManua.ls Logo

LINET AVE2 - Page 36

LINET AVE2
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D9U001PPB-010136/106
TABLE 1 – IMMUNITY to RF wireless communications equipment
TEST
FREQUENCY
(MHz)
BAND
(MHz)
SERVICE MODULATION
IMMUNITY
TEST LEVEL
(V/m)
385 380 - 390 TETRA 400
Pulse modulation
18 Hz
27
450 430 - 470
GMRS 460, FRS
460
FM ± 5 kHz
deviation
1 kHz sine
28
710
745
780
704 - 787
LTE band
13, 17
Pulse modulation
217 Hz
9
810
870
930
800 - 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Pulse modulation
18 Hz
28
1 720
1 845
1 970
1 700 - 1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE band
1, 3, 4, 25;
UMTS
Pulse modulation
217 Hz
28
2 450 2 400 - 2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE band 7
Pulse modulation
217 Hz
28
5 240
5 500
5 785
5 100 - 5 800
WLAN
802.11 a/n
Pulse modulation
217 Hz
9
NOTE There are applied no deviations to requirements of IEC 60601-1-2 ed. 4
NOTE There are no known other measures for keeping the basic safety based on EMC phenom-
ena