2.3 Electrical data
NOTE Upon request, LINET® can deliver critical care beds with electrical specifications that comply with
regional standards (custom voltage, different mains plugs).
Values
Input voltage
230 V, +/- 10%, 50 — 60 Hz
100 - 127V, +/- 10%, 50 — 60 Hz
Maximum power input
max. 1,6 A (input 230V), 370 VA
max. 3,7A (input 100–127V), 370VA
EN 60529 safety protection
IP 54
Safety class
Class I
Applied part
Type B
Duty cycle
10%, max. 2 min. out of 18 min.