Project SERVICE MANUAL
Date
05-2017
Version 05
83/84
9.2 Stop codes
Stop codes
Code
Reason
1
unexpected lateral tilt detected by main control box motherboard
accelerometer
2
foot control pressed after STOP
3
display button pressed after STOP
4
two buttons pushed simultaneously longer than 1 second
5
keyboard not released after system start
6
main control box temperature exceeded
7
column current overload 0
8
column current overload 1
9
column current overload 2
10
actuator current overload – backrest
11
actuator current overload – thighrest
12
stopped due to resistance on the same function
13
short circuiton on analog control (RC detection)
101
GO button not active
102
locked function
103
x-ray cassette inserted incorrectly
104
siderails folded down during lateral tilt
105
backrest angle greater than 30° during lateral tilt
106
lateral tilt stopped at 15° when using foot control
111
function blocked during one fault state
112
function blocked during active ALT
113
stopped due to risk of collision when bed is tilted
114
stopped in zero position
115
collision when exceeding threshold limit of calfrest and footrest
116
scale overload – tilt only to 15°
121
column potentiometer defect
122
stopped at 15° during lateral movement patient transfer
123
absolute scale overload
124
starting or running ALT on battery
125
Symbioso CPR (not stopping, only pop up)
See LINIS: page 1, STOP CODES

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET MULTICARE 2015 and is the answer not in the manual?

LINET MULTICARE 2015 Specifications

General IconGeneral
BrandLINET
ModelMULTICARE 2015
CategoryMedical Equipment
LanguageEnglish

Related product manuals