6.2.6 Positioning the weights for sensor 3 (left foot sensor)
6.2.7 Calculating and saving the new sensor constant - sensor 3:
sensor 3:
check the value on the secondary indicator
value of secondary display
x sensor 3 constant = new constant
value of weights used (80kg)
Write the new constant value
to the position of sensor 3
79,5
x 1.953 = 1,940
80