6.4 Switching between kg/lb – EU or US screen settings
Multicare 2015 redesigned beds can not be switched between kg and lb.Weighing in kg is set by default. Switching
between kg/lb is only allowed on beds for the U.S. market.
To allow switching between kg and lb:
Open the second page of the service menu (Sensor Sensitivity 2).
Press the kg/lb button, which is crossed out by default (see picture below).
It is possible to switch between weighing in kg or lb.
US version
EU version