Project SERVICE MANUAL
Date
05-2017
Version 05
73/84
7.4 Error codes
Code 1: System fault during standard mode
Code 2: System fault during ALT (Automatic lateral therapy)
7.5 Overheated main control box
If the temperature of the heat sink in the main control box reaches:
o 85°C /185°F – The battery charging function disconnects.
o 100°C / 212°F – The battery charging function disconnects, sound signalization occurs and
1 is added to LINIS (see LINIS chapter 7.1 – page 3)
o 115°C / 239°F – The bed stops functioning and sound signalization occurs.
The bed becomes functional again after the main control box cools down.
7.6 Power box overheating
If the transformer in the main control box reaches:
o 100°C / 212°F – The main control box automatically turns off and turns on automatically
after cooling.
o 130°C / 266°F – The heat fuse blows. It is necessary to replace the main control box.
7.7 One fault state
When the mains LED flashes, this signalises protection from one fault. In this state, only one bed
movement is possible and after this movement all functions will be blocked. This is indicated by all
lock LEDs flashing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET MULTICARE 2015 and is the answer not in the manual?

LINET MULTICARE 2015 Specifications

General IconGeneral
BrandLINET
ModelMULTICARE 2015
CategoryMedical Equipment
LanguageEnglish

Related product manuals