Project SERVICE MANUAL
Date
05-2017
Version 05
65/84
6.2.5 Calculating and saving the new sensor constant - sensor 2:
sensor 2:
divide the secondary indicator value by two
79,5 : 2 = 39,75
half of the value of the secondary display
x sensor 2 constant = new constant
half of the value of the weights used (40kg)
Write the new constant value
to the position of sensor 2
39.75
x 1.953 = 1,940
40