To calibrate the columns and accelerometers,
press and hold the CPR button until 3 cycles have
completed. This is possible from all control panels.
The bed will go into the lowest position and the
mattress platform parts will become horizontal.
After completing each cycle, a short sound
signalisation occurs and one lock LED turns off.
It is necessary to press the CPR button
throughout calibration otherwise the whole
process must be repeated!
Once the calibration cycles have completed, a
sound signalisation occurs and the lock LEDs turn
off. The calibration process is now finished.
It is recommended to quit service mode using the STOP button. Otherwise, service mode deactivates automatically
after approximately 3 mins.