Project SERVICE MANUAL
Date
05-2017
Version 05
49/84
Press the service mode button
Hold the button for at least 5 seconds until the
count down bar disappears and a sound signal
occurs.
To confirm that the bed is in service mode, the
green LED on the LCD flashes quickly.
To switch the RC circuit, press the leg rest lock
button
Short-circuit detection is switched ON when the
leg rest lock is not lit.
Short-circuit detection is switched OFF when the
leg rest lock is lit.
This mode is active for a duration of 5 mins.