Project SERVICE MANUAL
Date
05-2017
Version 05
57/84
Enter the 5-digit code which is composed of the
last numbers of the date and time.
Press the green enter button.
4. 5. 1. 2. 3. (example 49011)
Scale calibration is always needed when:
o the scale module is replaced.
o any of the tensometers is replaced.
The gravitational constant, according to the bed‘s location and values (voltage) of the tensometers, must be
set and then the bed calibrated when replacing the scale module.
The values (voltage) of the replaced tensometers must be set and then the bed calibrated when replacing a
tensometer.