Project SERVICE MANUAL
Date
05-2017
59/84
Press the button for scale calibration.
Hold the button for at least 5 seconds until the
count down bar disappears and a sound signal
occurs.
It is necessary to perform each step according to the manual when calibrating the scale:
Factory zero
Scale calibration at 160 kg (alternatively 80kg)
Saving the calibration constant to memory.