Page #40 background imageLoading...
Page #40 background image
D9U003VB3-0101_0440/65
10.2 AccumulatorBack-upPack
SCUs with the back-up accumulator feature can be equipped with an optional accumulator back-up pack.

Accumulator Status Indi-
cator
Meaning Visual Signal and Audio Alert Signal
POWER FAIL MODE
or
Accumulator TOO LOW TO USE
1 status of Accumulator Status Indicator
Power Fail Alert (2-tone alert, sounds until bac-
kup memory is erased):
when the backup accumulator is
󰀨
mains power)
when SCU is running from accu-
mulator if the accumulator is too
low to use
CHARGE FAULT
or
Accumulator FAULT
(SCU is connected to the mains)
2 statuses of Accumulator Status Indicator in
repetitive sequence (for 5 seconds not lit LEDs

no audio alert
Accumulator CORRECTLY FITTED
(SCU is connected to the mains)
2 statuses of Accumulator Status Indicator in
sequence (LEDs are lit for 2 seconds)
2 seconds long beep
To mute the audio alert signal:
Mute audio alert signal (see „Mute“ - chapter 11.2.2).
Fig. Identication label of the accumulator back-up pack