12 Cleaning/Disinfection
For safe and gentle cleaning:
Disconnect SCU from mains.
Do not use any strong acids or bases (optimum pH range 6 - 8).
Only use detergents that are suitable for cleaning medical equipment.
Do not use abrasive powders, steel wool, or other material and cleaning agents that might damage the
mattress replacement system.
Never use any corrosive or caustic detergents.
Never use detergents that deposit calcium carbonate.
toluene, acetone, etc.).
Clean electrical components carefully and allow them to dry fully. It is recommended to use the cleaning
wipes.
Neither immerse SCU in water nor heat or steam-clean it.
Observe local directives concerning infection control.
LINET®recommendsthefollowingcleaningagents:
12.1 Cleaning
CAUTION!
Incorrect cleaning/disinfection can damage the mattress and SCU!
Do not use pressure or steam cleaners.
Use the recommended cleaning agents only.
Follow the instructions and observe the dosages recommended by the manufacturer.
cleaning.
Cleaning agents Manufacturer
Presept Johnson&Johnson
Terralin, Mikrozid, Thermosept Schülke&Mayr
Bacillol Plus, Bacillocid Rasant, Microbac Forte, Dismozon Pur BODE Chemie
Type of Cleaning Parts to Clean
Routine Cleaning and Disinfection exposed mattress parts
exposed SCU parts
Full Cleaning and Disinfection exposed mattress parts
exposed SCU parts
internal parts of mattress
internal parts of cover