- 37 -
Serial protocol alarms:






MODE
Bit LSB
76543210
xxxxxxx1
76543210
xxxx1xxx
76543210
xxxxxx1x
76543210
xxxxx1xx
76543210
On gross:
xxx1xxxx
On net:
xx1xxxxx
The response to the
zero command is a
value not valid” error
(error code 3)
Status
Register
MODBUS RTU
ASCII
__O-F_
__O-L_
__O-F_
__O-L_
__O-F_
&aa#CR
RIP *
__O-F_
__O-L_
__O-F_
__O-L_
__O-F_
__O-F_
HDRIP-N
_ERCEL
_ER_OL
_ER_AD
######
_ER_OF
O__SET
CONTIN
_ERCEL
_ER_OL
_ER_AD
^^^^^^
_ER_OF
O__SET
* For RIP remote displays, if the message exceeds 5 digits the display reads .
With an alarm the relays open and the analog outputs go to the lowest possible value
according to the following table:
RANGE
0÷20 mA
4÷20 mA
0÷5 V
0÷10 V
±10 V
±5 V
Output value
-0.2 mA
3.5 mA
-0.5 V
-0.5 V
0 V
0 V