Page | 23
4.
Set the Span Frequency to
500 kHz
5.
The 80 MHz standard signal of the inside of the analyzer is about -
40 dBm
, so
using the scroll knob, step keys, or the numeric keypad, adjust the reference
level to
-30 dBm.
6.
Set the marker by pressing the MARKER key and moving the marker to the
maximum value of the signal, or by pressing PEAK SEARCH.
Please read the amplitude and frequency. It should be indicated as in figure 3-5. The
frequency is indica
ted horizontally, and the amplitude vertically.
[Figure 3-5] Analyzed Results of Amplitude and Frequency