If you press Meas Set-up after selecting Occupied BW, you can set the occupied
bandwidth power measurement parameter. If you press Meas Control after
selecting Occupied BW, you can temporarily suspend or resume measurement or
make conversion between continuous or single sweep modes.
MEASURE >> Occupied BW
5-12. Preset (Reset)
The PRESET key initializes the analyzer back to the factory default state. As for the set-up of the
preset state, please refer to the following table of conditions. If you wish to select Preset (Factory),
press SYSTEM >> Power On/Preset >> Preset Type (Factory).
[Table 5-1] Factory Preset Conditions
Amplitude Unit
dBm
Indication and Lattice Display
On
Attenuation
30 dB
Center Frequency
1.5 GHz
Start Frequency
0 Hz
Stop Frequency
3.0 GHz
CF Step
10% of a span
Detect
Normal
Display Line
-25 dBm, display turns off.
Frequency Offset
0 Hz
Log Scale
10 dB per sector
Reference Level
0 dBm
Reference Level Offset
0 dB
Marker
Off
Resolution Bandwidth
3 MHz ( Auto )
Video Bandwidth
3 MHz ( Auto )
VBW/RBW Ratio
1.000
Video Avg.
Off
Span
3.0 GHz