Page | 65
sweep starts from the ascending/descending angle of the signal wave form inputted
from the EXT TRIG input connector on the back panel. To start the trigger, a TTL
input signal is required.
TRIG >> External
■
Trig Delay On Off
You can set the delay of the time that the analyzer is waiting, or turn it on or off
before starting a sweep after receiving the external trigger signal. This function can
be used only when there is an external trigger.
TRIG >> Trig Delay
5-22. Trace
Allows the user to manipulate and save trace information displayed on the LCD screen. The
spectrum analyzer renews the trace information after every sweep.
■
Trace 1 2 3
To select a trace, press the soft key Trace 1 2 3 until the number of the desired
trace is underlined.
TRACE >> Trace
■
Clear Write
This deletes data previously saved in the trace memory. It saves amplitude data
newly acquired after sweeping, and indicates the amplitude data on the LCD display
screen. This function is applied to trace 1 when inputting power.
TRACE >> Clear Write
■
Max Hold
The Max Hold function displays and continuously updates the trace to display the
highest peak amplitude of signal level.
TRACE >> Max Hold