Model 3030 & 3030E Series Technical Manual Section 6
Ludlum Measurements, Inc. Page 16 November 2017
Press the COUNT button. When count cycle is complete, multiply the
β BETA LCD reading by 10 and confirm that this reading is within
±2% of the incoming count rate. Repeat this step with 200 mV
pulses and check the α-ALPHA LCD reading.
The objective of the next steps is to find an operating voltage for the
detector that will yield the detector’s stated efficiency while not over-
counting the background radiation field. The 2π efficiency (Eff) is
calculated as:
Instrument Count Rate
×100% = Eff
cpm of Source
In addition, cross talk must be considered for determining
the optimum operating voltage and is defined as:
Alpha cross talk =
αSource - BackCβ
×100%
αNet
and
Beta cross talk =
βSource BackCα
×100%
βNet
where
αSource = α source count in β channel
BackCβ = background count in β channel
αNetCα = Net α source count in α channel
βSource =β source count in α channel
BackCα = background count in α channel
βNetCβ = Net β source count in β channel
For the alpha/beta scintillation detector, applicable 4π efficiencies
are: