Model 3030 & 3030E Series Technical Manual Section 7
Ludlum Measurements, Inc. Page 38 November 2017
tttt.t = count time in minutes (If count is in progress, this value
shows count time remaining.)
RS – Read Samples
Reads all logged samples from memory. While reading samples, the audio is
stopped and the displays show a scrolling series of dashes along with the
sample number. A "$" marks the end of the samples.
Returns: ssss MM/DD/YY HH:MM:SS aaaaaa bbbbbb ttttt.t x
cccccccccc[CR][LF] - 59 bytes $[CR][LF] - 3 bytes
ssss = sample number
aaaaaa = alpha scaler
bbbbbb = beta scaler
x = S (Scaler), C (cpm), and D (dpm)
ttttt.t = count time in seconds
cccccccccc = user defined comment
RT – Read Current Count Time
Reads the current count time in seconds. Returns nnnnn.n.
RU – Read Comment
Reads the user-defined, 10-character comment field. Returns xxxxxxxxxx.
RV – Read High Voltage
Reads the high voltage set point. Returns nnnn.
RXn – Read the Crosstalk
Reads the beta-to-alpha or alpha-to-beta crosstalk. The value must be
divided by 10 to get the correct value. Returns nnn.
1 = Read beta-to-alpha crosstalk.
2 = Read alpha-to-beta crosstalk.
RY – Read Recycle Mode Status
Reads the status of the recycle mode. Returns OFF or ON.
Rn – Read the Alarm Set Points
Reads the specified alarm set point. Returns nnnnnn.