Model 3030 & 3030E Series Technical Manual Section 7
Ludlum Measurements, Inc. Page 38 November 2017
tttt.t = count time in minutes (If count is in progress, this value
shows count time remaining.)
RS Read Samples
Reads all logged samples from memory. While reading samples, the audio is
stopped and the displays show a scrolling series of dashes along with the
sample number. A "$" marks the end of the samples.
Returns: ssss MM/DD/YY HH:MM:SS aaaaaa bbbbbb ttttt.t x
cccccccccc[CR][LF] - 59 bytes $[CR][LF] - 3 bytes
ssss = sample number
aaaaaa = alpha scaler
bbbbbb = beta scaler
x = S (Scaler), C (cpm), and D (dpm)
ttttt.t = count time in seconds
cccccccccc = user defined comment
RT Read Current Count Time
Reads the current count time in seconds. Returns nnnnn.n.
RU Read Comment
Reads the user-defined, 10-character comment field. Returns xxxxxxxxxx.
RV Read High Voltage
Reads the high voltage set point. Returns nnnn.
RXn Read the Crosstalk
Reads the beta-to-alpha or alpha-to-beta crosstalk. The value must be
divided by 10 to get the correct value. Returns nnn.
1 = Read beta-to-alpha crosstalk.
2 = Read alpha-to-beta crosstalk.
RY Read Recycle Mode Status
Reads the status of the recycle mode. Returns OFF or ON.
Rn Read the Alarm Set Points
Reads the specified alarm set point. Returns nnnnnn.