Model 3030 & 3030E Series Technical Manual Section 7
Ludlum Measurements, Inc. Page 39 November 2017
1 = Read alpha alarm set point.
2 = Read beta alarm set point.
3 = Read alpha + beta alarm set point.
Xn – Set Crosstalk Mode
Sets crosstalk ON/OFF or return crosstalk status, where “n” is a value of 0,
1, or 2.
0 = Sets crosstalk OFF.
1 = Sets crosstalk ON.
2 = Returns status.
SBnxxxxxx – Set Background Subtract
Sets the alpha or beta background subtract.
1 = Set alpha background subtract.
2 = Set beta background subtract.
SCmmddyyyy – Set Calibration Date
Sets the calibration due date.
SDmmddyyyyz – Set Date
Sets the date on the read time clock. The “z” is the value of the year Mod 4.
SEnxxx – Set Efficiency
Sets the alpha or beta efficiency. The efficiency must be multiplied by 10
before sending.
1 = Set alpha efficiency.
2 = Set beta efficiency.
SHnxxxxxxxxxxxxxxx – Set Header
Sets a user-defined header. There are six headers accessed 1-6. Each header
is 15 characters long.
SLyyyyy.y – Set Loss of Count Timer
Sets the loss of count timer in seconds. The max is 99999.9 seconds.
SNnyyyyyy – Set MDA
Sets the alpha and beta MDA.