Err.CAL
Calibration parameters lost – send the transducer
to a service, the message does not prevent measured
values from being displayed, message is displayed
incycles.
Err Batt
Service
Realtimeclockbatterylowvoltage–lossofrealtimeclock
presets after a power loss, the transducer can operate,
consider sending the transducer to a service to replace
the battery; the message does not prevent measured
valuesfrombeingdisplayed,messageisdisplayedincycles.
Changingdateorhoursettingsswitchesofthatmessage.
Err.PAR
Parameter error – restore default settings, do not ope-
rate the transducer until default settings are restored,
themessagedoesnotpreventmeasuredvaluesfrombeing
displayed,messageisdisplayedincycles.
Err Ind
Wrongconfigurationofindividualcharacteristicparameters,
thetransducercanoperate–individualcharacteristicfun-
ctiondoesnotwork,themessage does notprevent mea-
sured values from beingdisplayed, messageis displayed
incycles.
Error
Par.File
Reading configuration from file stored on an external
SD/SDHC card or on the internal file system memory
unsuccessful– file ismissing orcorruped, thetransducer
can beoperated,themessagedoesnotpreventmeasured
valuesfrombeingdisplayed,messageisdisplayedincyc-
lesforabout20seconds.