5-11 Data Logger Function
Remark :
Please do the following , then execution data logger function.
1. First connect the mains adaptor, with no batteries installed.
2. Check that the unit power on.
3. then install the batteries, as a backup in case there is a power
failure during data logging.
4. After installing the battery, remove the adaptor and confirm that the
meter action is normal after installing the battery. If no problem,
connect the adaptor to the meter.
5. Then start data logging.
A : Press REC KEY once to start the data record function.
A-1 : If the bottom right shows as " Change Card ", it
indicates the memory space is already full either or the
SD CARD exist some wrong.
A-2 : If the SD CARD is normal, the data logger function
will start to be executed.
V12:
0.0 V
V1:
0.0 V
A1:
0.00 A
V23:
0.0 V
V2:
0.0 V
A2:
0.00 A
V31:
0.0 V
V3:
0.0 V
A3:
0.00 A
P1:
-0.000 KW
S1:
0.000 KVA
Q1:
- 0.000 KVAR
P2:
-0.000 KW
S2:
0.000 KVA
Q2:
- 0.000 KVAR
P3:
-0.000 KW
S3:
0.000 KVA
Q3:
- 0.000 KVAR
PΣ :
-0.000 KW
SΣ:
0.000 KVA
QΣ:
- 0.000 KVAR
PF1:
- 0.00
PF2:
- 0.00
PF3:
- 0.00
PFΣ:
0.00
PFH:
0.00
Φ1:
- 0.0°
Φ2:
- 0.0°
Φ3:
- 0.0°
WH:
0.000 KWH
SH:
0.000 KVAH
QH:
0.000 KVARH
FREQ:
0.0 Hz
CP1201 Change
20A 4W SEC: 2 CT: 1 PT: 1 Card
B : The bottom right display will show the recorded data points.
B-1 : Each file can record up to 30,000 data points as
screen 1 when the record points exceed 30,000 points,
system will create a new file automatically. (For example,
WTA01001.XLS will be replaced by WTA01002.XLS)
B-2 : While pressing REC KEY once, the data logger
function will stop to execute, the record points will
disappear on bottom right display as screen 2.
63