EasyManua.ls Logo

MagCore Plus II - Program Update; Clean the Waste Box

MagCore Plus II
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
MagCore® Plus II
Remove back lid and check the gap at the back of the machine. There should be no
gap!
If the rack is pushed all the way but there is still a gap, adjust the position of the
metal plates on the cartridge rack by loosening 2 screws and moving the plate
forward or backward:
3. If the problem remains, adjust X home sensor block plate, as shown in the picture
below:
-Loosen 2 screws (circled in red)
-Move block plate back and forth, as indicated by the arrow
-Tighten the screws back
Recalibrate X-value following the calibration procedure (p.12)

Other manuals for MagCore Plus II