Operation Manual easyScreen
Frequency response of the ear canal with
click stimulus
RMS measurement in frequency domain,
average of frequency components around
the DP frequency (26 bins < 2500 Hz and
60 bins ≥ 2500 Hz)
View of progress bars toward Pass,
feedback about Noise, recording Time
Depends on protocol
Default: Minimum 6 dB
Depends on protocol
Default: Minimum -5 dB SPL
Depends on selected protocol
Default: 3 of 4
Non-Linear click (according to IEC 60645-3)
60 dB peSPL to 83 dB peSPL
83 dB peSPL (peak to peak calibrated), auto
in-ear calibration
Default center
frequencies
1400, 2000, 2800, 4000 Hz
View of progress bars toward Pass,
feedback about Probe stability, Noise and
recording Time
Depends on protocol
Default: Minimum 4 dB
Depends on protocol
Default: Minimum -5 dB SPL
Depends on protocol
Default: 3 of 4