5.2.3 Conditioning tubes
Tubes will require conditioning when freshly
packed and when they have been stored for
longer than 24 hours (for trace level analysis)
or 7 days (for ppm level analysis), without
being fitted with ¼-inch brass storage caps.
Tubes should be cleaned using the dedicated
tube conditioning mode on UNITY or using the
TC-20 Multi tube conditioning rig.
See Application information (TDTS 5 /
TDTS19) for further details.