EasyManua.ls Logo

Masimo SET Radical-7 - Page 24

Masimo SET Radical-7
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
6-4
Symptom Cause and Corrective Action
Steps
Download Procedure.
2. Replace The Interface Cable
3. Replace the Portable System
Board
4. Replace the Docking Station
System Board
No LED'S Illuminated
1. Both Portable and Docking
Station are turned on.
2. Refer to Radical-7
Operators
Manual for LED status
Parameters.
Power fault
1. Replace external Power Cable
2. Check fuses at AC Input Filter
3. Check 12VDC power (Sections 7
& 8)
4. Replace Docking Station System
Board
No Battery Icon Displayed NiMH Battery Fault
1. Replace NiMH battery
2. Replace Portable System Board
Battery LED Not
Illuminated
1. No fault, battery is fully charged
2. Check 12VDC power (Sections 7
& 8)
3. Replace Docking Station System
Board
Battery LED Flashing
1. Ensure Service Mode – Deep
Discharge is not enabled
2. Allow NiMH Battery to cool down
3. Verify Fan operation
4. Replace NiMH Battery
5. Replace Docking Station System
Board.
Alarm LED does not
Illuminate
1. Unit must be in alarm state
2. Audible alarm can be heard
1. Replace Docking Station System
Board
2. Replace LED Cable.
Alarm Condition But No
Audible Indication.
1. Unit must be in alarm
condition.
2. Should hear sound upon
button depression.
Alarm Fault
1. Replace Speaker
2. Replace Portable System Board
Display Missing Segments
1. Replace LCD
2. Replace Portable System Board
Display Backlight Will Not
Illuminate
1. Replace Portable System Board
2. Replace Display
Button Function Error
1. Replace Keypad

Related product manuals