EasyManua.ls Logo

Masimo NomoLine ISA AX+ - Page 27

Masimo NomoLine ISA AX+
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NomoLine ISA Service Manual Chapter 2: Maintenance Procedures
www.masimo.com 25 Masimo
Item
Description Item Description
2 Leave
disconnected
4 Connect to the ISA Input Connector.*
* See NomoLine ISA CO2/AX+ Interface on page 11 or ISA OR+ Interface on page 12.
Procedure
1. Configure the HM28 device to display measurement values in hPa.
2. Press "MODE" button until the display shows MAX/MIN values as in figure 1-13.
Figure 1-13
3. Make sure the NomoLine ISA system is powered ON.
4. Connect the NomoLine ISA Maintenance Kit to the NomoLine ISA Input Connector
shown in figure 1-12 and leave the exhaust open. See NomoLine ISA CO2/AX+
Interface on page 11 or ISA OR+ Interface on page 12.
Note: Verify the 4-way stopcock lever position in figure 1-12.
5. Press "CLEAR/ZERO" button on the HM28 digital manometer.
6. Wait for 30 seconds.
7. Calculate a mean value based on the MAX/MIN values displayed on the HM28
digital manometer.
Test result:
The NomoLine ISA system sampling flow rate
1
is inside specified range if the mean value is
within specified acceptance criteria
2
. The NomoLine ISA system should not be used if it fails
to meet the acceptance criteria limits.
1
Sampling flow rate: 50± 10 sml/min
2
Acceptance criteria: 16 to 24 hPa
Gas Span Check
Overview
The gas span check procedure requires the use of a suitable calibration gas mixture. A
cylinder containing such mixture may be acquired from the vendor of your choice, provided
the following gas concentrations and accuracy are met.

Related product manuals