Root with Pathway Operation
www.masimo.com 14 Masimo
Options Description Factory Default
Settings
User Configurable
Settings
Alarm
Delay
When an alarm condition is met, this
feature delays the audible part of an
alarm.
10 0, 5, 10, or 15
SpO2 Target Range
Use the SpO2 Target Range settings screen to configure the Glide Path visualization. See
Pathway Screen on page 20.
Option Description Factory Default Settings* User Configurable Settings**
1 Min
SpO
2
Max 60 60 to 98 in steps of 1
SpO
2
Min 65 61 to 99 in steps of 5
2 Min
SpO
2
Max 65 60 to 98 in steps of 1
SpO
2
Min 70 61 to 99 in steps of 1
3 Min
SpO
2
Max 70 60 to 98 in steps of 1
SpO
2
-axis Min 75 61 to 99 in steps of 1
4 Min
SpO
2
Max 75 60 to 98 in steps of 1
SpO
2
Min 80 61 to 99 in steps of 1
5 Min
SpO
2
Max 80 60 to 98 in steps of 1
SpO
2
Min 85 61 to 99 in steps of 1
10 Min
SpO
2
Max 85 60 to 98 in steps of 1
SpO
2
-axis Min 95 61 to 99 in steps of 1
* Default settings are established based upon the recommendation made by Wyckoff MH, Aziz
K, Escobedo MB, Kapadia VS, Kattwinkel J, Perlman JM, Simon WM, Weiner GM, Zaichkin, JG.
Part 13: neonatal resuscitation: 2015 American Heart Association Guidelines Update for

Other manuals for Masimo Root

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo Root and is the answer not in the manual?

Masimo Root Specifications

General IconGeneral
CategoryMedical Equipment
Display10.1 inch touchscreen
ConnectivityWiFi, Bluetooth
Power SupplyAC power
Dimensions26.8 cm x 22.4 cm x 5.1 cm
PortsUSB, Ethernet
Supported ParametersSpO2, Pulse Rate
CompatibilityMasimo sensors
ModelRoot

Related product manuals