Root with Pathway Operation
www.masimo.com 15 Masimo
Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation.
2015;132(suppl 2):S543–S560.
** Visual indicators only. Audible alarms only occur when the Radical-7 high and low alarm
limits are reached. Refer to the Operator's Manual, Radical-7® Pulse CO-Oximeter®.
PR Threshold
From the PR Threshold settings screen, change any of the following options. Default settings
are established based upon the recommendation made by the International Liaison
Committee on Resuscitation (ILCOR).
Options Description Factory Default
Settings*
User Configurable
Settings**
Low Limit Adjusts the low limit threshold
visual indicator.
100 35 to 235 in steps of 5
Critical
Limit
Adjusts the critical limit threshold
visual indicator.
60 30 to 230 in steps of 5
* Default settings are established based upon the recommendation made by Wyckoff MH, Aziz
K, Escobedo MB, Kapadia VS, Kattwinkel J, Perlman JM, Simon WM, Weiner GM, Zaichkin, JG.
Part 13: neonatal resuscitation: 2015 American Heart Association Guidelines Update for
Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation.
2015;132(suppl 2):S543–S560.
** Visual indicator only. Audible alarms only occur when the Radical-7 high and low alarm
limits are reached. Refer to the Operator's Manual, Radical-7® Pulse CO-Oximeter®.
Device Settings
On Root equipped with Pathway, some Device Settings options change and are reflected
below:
Access Control
See Access Control on page 16
.