Root with Pathway Pathway
www.masimo.com 26 Masimo
PR Visual Indicators
PR thresholds are configurable in the PR Threshold settings menu. See PR Threshold on page
15.
PR Low (Below 100 bpm)
The PR trend and current indicator display yellow after breaching the 100 bpm dashed limit.
A visual indicator displays immediately.
• Yellow glow from the bottom of the PR trend.
• No message displays in the header.
• No audible alarms.
PR Critical (Below 60 bpm)
The PR trend and current indicator display red after breaching the 60 bpm solid limit. A
visual indicator displays immediately.
• Red glow from the bottom of the PR trend.
• No message displays in the header.
• No audible alarms.
Time Adjustment
Time adjustment allows the measurement start time to be adjusted in relation to actual birth
time.
• Time adjustment can only be initiated within the first ten minutes of a session.
• No time adjustment can be made to less than 0 minutes.