3
rd
load set-point (Disp, Occurrence)
4
th
load set-point (Disp, Occurrence)
5
th
load set-point (Disp, Occurrence)
.load at time (Time)
Obviously no occurrence parameter is needed for this command, because
time can only move in one direction!
1
st
displacement set-point (Load, Occurrence)
2
nd
displacement set-point (Load,
Occurrence)
3
rd
displacement set-point (Load,
Occurrence)
4
th
displacement set-point (Load, Occurrence)
5
th
displacement set-point (Load, Occurrence)
displacement at time (Time)
Obviously no occurrence parameter is needed for this command, because
time can only move in one direction!
stiffness (Disp1, Disp2)
The software will calculate the stiffness of the specimen between Disp1, and
Disp2, by looking up the loads (L
1
and L
2
) corresponding to the first
occurrence of these two displacements (D
1
and D
2
) and returning the value
of:
(L
1
- L
2
) / (D
1
- D
2
)
in the context of spring testing this would be called a ‘spring rate’ calculation.
Pressing the [1] key will toggle between ‘ON’
& ‘OFF’. If this calculation is active the difference between the ‘absolute zero