Mecmesin MultiTest-x manual
23
• 3
rd
load set-point (Disp, Occurrence)
•
• 4
th
load set-point (Disp, Occurrence)
• 5
th
load set-point (Disp, Occurrence)
• .load at time (Time)
Obviously no occurrence parameter is needed for this command, because
time can only move in one direction!
• 1
st
displacement set-point (Load, Occurrence)
• 2
nd
displacement set-point (Load,
Occurrence)
• 3
rd
displacement set-point (Load,
Occurrence)
• 4
th
displacement set-point (Load, Occurrence)
• 5
th
displacement set-point (Load, Occurrence)
• displacement at time (Time)
Obviously no occurrence parameter is needed for this command, because
time can only move in one direction!
• stiffness (Disp1, Disp2)
The software will calculate the stiffness of the specimen between Disp1, and
Disp2, by looking up the loads (L
1
and L
2
) corresponding to the first
occurrence of these two displacements (D
1
and D
2
) and returning the value
of:
(L
1
- L
2
) / (D
1
- D
2
)
in the context of spring testing this would be called a ‘spring rate’ calculation.
• Pressing the [1] key will toggle between ‘ON’
& ‘OFF’. If this calculation is active the difference between the ‘absolute zero