17 | Initial startup
Checks before use
– Check Thopaz system before use for damage of the power cord or plug, obvious
device damage or safety defects and proper functioning of the device.
– Check for completeness and general condition of Thopaz delivery package.
7.2a
3. 3.1 Charge for approximately
3 hours or until the [ ]
symbol appears in the
display.
3.2 Press [
] to switch
Thopaz on. A selftest
starts.
Important: Do not yet connect the patient to the system.
2. 2.1 Connect Thopaz to mains
by mains adapter
077.0148 or by docking
station 079.0037.
Initial startup
procedure
1. 1.1 Select plug.
1.2 Attach plug to the mains
adapter
077.0148.
7.2b
Initial startup
7.2
WARNINGS
For use only after instruction by medically trained and qualified persons who have been
adequately trained in the use of Thopaz.
CAUTIONS
Before the first application in accordance with the intended use, an inspection
according to IEC 62353 (see Appendix B) is recommended.